Back to Search Start Over

Two to one CRTD pacing: When the T wave and the P wave interact

Authors :
Yuval Konstantino
Aref Alnsasra
Hilmi Alnsasra
Moti Haim
Source :
Journal of electrocardiology. 50(3)
Publication Year :
2016

Abstract

We presented a unique phenomenon of 2:1 cardiac resynchronization therapy pacing due to T wave oversensing. Ultimately, by utilizing a unique feature of integrated bipolar sensing, we succeeded to eliminate the T wave oversensing signals, and restore 1:1 CRTD pacing. Importantly, this feature enabled us to overcome the T wave oversensing issue, without the need to decrease the ventricular sensitivity, which could potentially interfere with ventricular arrhythmia detection and appropriate shock delivery when required.

Details

ISSN :
15328430
Volume :
50
Issue :
3
Database :
OpenAIRE
Journal :
Journal of electrocardiology
Accession number :
edsair.doi.dedup.....41f1706956fb0f282002bf30ac281c91