Simultaneous multi‐slice cardiac cine with Fourier‐encoded self‐calibration at 7 Tesla
- Source :
- Magnetic Resonance in Medicine, Magnetic Resonance in Medicine, 2018, 81 (4), pp.2576-2587. ⟨10.1002/mrm.27593⟩, Magnetic Resonance in Medicine, Wiley, 2018, 81 (4), pp.2576-2587. ⟨10.1002/mrm.27593⟩
- Publication Year :
- 2018
- Publisher :
- HAL CCSD, 2018.
-
Abstract
- International audience; PurposeTo accelerate cardiac cine at 7 tesla using simultaneous multi‐slice (SMS) acquisition with self‐calibration to resolve misalignment between calibration and imaging data due to breathing motion.MethodsA spoiled‐gradient echo cine sequence was modified with radiofrequency phase‐cycled SMS excitations. A Fourier encoding strategy was applied along the cardiac phase dimension to allow for slice untangling and split‐slice GRAPPA calibration. Split‐slice GRAPPA was coupled with regular GRAPPA (SMS‐GRAPPA) and L1‐SPIRiT (SMS‐L1SPIRiT) for image reconstruction. 3‐slice SMS cine MRI was evaluated in ten subjects against single‐slice cine MRI in terms of SNR and contrast‐to‐noise ratio and slice leakage.ResultsSNR decreased significantly from 10.1 ± 7.1 for single‐slice cine to 7.4 ± 2.8 for SMS‐GRAPPA (P = 0.02) and was recovered to 9.0 ± 4.5 with SMS‐L1SPIRiT (P = 0.02). Contrast to noise ratio decreased significantly from 14.5 ± 8.1 for single‐slice cine to 5.6 ± 3.6 for SMS‐GRAPPA (P < 0.0001) and increased slightly but significantly back to 6.7 ± 4.4 for SMS‐L1SPIRiT (P = 0.03). Specific absorption rate restrictions imposed a reduced nominal flip angle (−37 ± 7%, P = 0.02) for 3‐slice SMS excitations compared to single‐slice acquisitions. SMS slice leakage increased significantly from apex (8.6 ± 6.5 %) to base (13.1 ± 4.1 %, P = 0.03) in the left ventricle.ConclusionThree‐fold acceleration of cine at 7T was achieved using the proposed SMS technique. Fourier encoding self‐calibration and regularized image reconstruction enabled simultaneous acquisition of three slices without significant SNR decrease but significant CNR decrease linked to the reduced nominal excitation flip angle.
- Subjects :
- Adult
Male
Time Factors
[SDV.IB.IMA]Life Sciences [q-bio]/Bioengineering/Imaging
Fourier encoding
Magnetic Resonance Imaging, Cine
Iterative reconstruction
Signal-To-Noise Ratio
Simultaneous Multi-Slice (SMS)
030218 nuclear medicine & medical imaging
Motion
Young Adult
03 medical and health sciences
symbols.namesake
0302 clinical medicine
Nuclear magnetic resonance
Contrast-to-noise ratio
Flip angle
Cardiac function MRI (cine)
Image Interpretation, Computer-Assisted
Image Processing, Computer-Assisted
Calibration
Humans
Radiology, Nuclear Medicine and imaging
Physics
Fourier Analysis
Echo-Planar Imaging
Respiration
Specific absorption rate
Heart
Multi slice
ultra-high field MRI
Image Enhancement
Healthy Volunteers
Fourier transform
7 Tesla cardiac magnetic resonance (CMR)
symbols
cardiovascular system
[PHYS.PHYS.PHYS-MED-PH]Physics [physics]/Physics [physics]/Medical Physics [physics.med-ph]
Female
Artifacts
030217 neurology & neurosurgery
Cardiac phase
Subjects
Details
- Language :
- English
- ISSN :
- 07403194 and 15222594
- Database :
- OpenAIRE
- Journal :
- Magnetic Resonance in Medicine, Magnetic Resonance in Medicine, 2018, 81 (4), pp.2576-2587. ⟨10.1002/mrm.27593⟩, Magnetic Resonance in Medicine, Wiley, 2018, 81 (4), pp.2576-2587. ⟨10.1002/mrm.27593⟩
- Accession number :
- edsair.doi.dedup.....3f8eb5592f25bd268ada540de8e5ec32
- Full Text :
- https://doi.org/10.1002/mrm.27593⟩