A Recirculation Aerosol Wind Tunnel for Evaluating Aerosol Samplers and Measuring Particle Penetration through Protective Clothing Materials

Authors :
Ta Chih Hsiao
Pengfei Gao
Peter A. Jaques
Source :
The Annals of Occupational Hygiene.
Publication Year :
2011
Publisher :
Oxford University Press (OUP), 2011.

Abstract

A recirculation aerosol wind tunnel was designed to maintain a uniform airflow and stable aerosol size distribution for evaluating aerosol sampler performance and determining particle penetration through protective clothing materials. The oval-shaped wind tunnel was designed to be small enough to fit onto a lab bench, have optimized dimensions for uniformity in wind speed and particle size distributions, sufficient mixing for even distribution of particles, and minimum particle losses. Performance evaluation demonstrates a relatively high level of spatial uniformity, with a coefficient of variation of 1.5-6.2% for wind velocities between 0.4 and 2.8 m s(-1) and, in this range, 0.8-8.5% for particles between 50 and 450 nm. Aerosol concentration stabilized within the first 5-20 min with, approximately, a count median diameter of 135 nm and geometric standard deviation of 2.20. Negligible agglomerate growth and particle loss are suggested. The recirculation design appears to result in unique features as needed for our research.

Details

ISSN :
14753162
Database :
OpenAIRE
Journal :
The Annals of Occupational Hygiene
Accession number :
edsair.doi.dedup.....385f7479dd6cbd9f8135a6497831ca1d
Full Text :
https://doi.org/10.1093/annhyg/mer062