Back to Search
Start Over
Outflow graft foreign body reaction and thrombosis in HeartMate 3 left ventricular assist device
- Source :
- Asian Cardiovascular and Thoracic Annals. 30:596-600
- Publication Year :
- 2021
- Publisher :
- SAGE Publications, 2021.
-
Abstract
- Outflow cannula occlusion is an infrequent complication occurring among recipients of continuous flow left ventricular assist devices. Hereby, we present a case of a 66-year-old man with a HeartMate 3 left ventricular assist device that presented 18 months after implantation and during his admission, evidence of poor left ventricular unloading and forward flow uncovered intrinsic and extrinsic outflow cannula obstruction. The patient’s course was complicated by cardiogenic shock, diffuse alveolar hemorrhage and multiorgan failure. Post-mortem examination suggested foreign body reaction as a potential contributor of outflow graft obstruction at the level of the bend relief.
- Subjects :
- Male
Pulmonary and Respiratory Medicine
medicine.medical_specialty
medicine.medical_treatment
Internal medicine
Occlusion
medicine
Humans
Aged
Heart Failure
business.industry
Foreign-Body Reaction
Cardiogenic shock
Thrombosis
Diffuse alveolar hemorrhage
General Medicine
medicine.disease
Cannula
Treatment Outcome
Ventricular assist device
cardiovascular system
Cardiology
Surgery
Outflow
Heart-Assist Devices
Foreign body
Cardiology and Cardiovascular Medicine
business
Subjects
Details
- ISSN :
- 18165370 and 02184923
- Volume :
- 30
- Database :
- OpenAIRE
- Journal :
- Asian Cardiovascular and Thoracic Annals
- Accession number :