Back to Search
Start Over
Is There a Difference in Screw Accuracy, Robot Time Per Screw, Robot Abandonment, and Radiation Exposure Between the Mazor X and the Renaissance? A Propensity-Matched Analysis of 1179 Robot-Assisted Screws
- Source :
- Global spine journal.
- Publication Year :
- 2021
-
Abstract
- Study Design: Prospective single-cohort analysis. Objectives: To compare the outcomes/complications of 2 robotic systems for spine surgery. Methods: Adult patients (≥18-years-old) who underwent robot-assisted spine surgery from 2016-2019 were assessed. A propensity score matching (PSM) algorithm was used to match Mazor X to Renaissance cases. Preoperative CT scan for planning and an intraoperative O-arm for screw evaluation were preformed. Outcomes included screw accuracy, robot time/screw, robot abandonment, and radiation. Screw accuracy was measured using Vitrea Core software by 2 orthopedic surgeons. Screw breach was measured according to the Gertzbein/Robbins classification. Results: After PSA, a total of 65 patients (Renaissance: 22 vs. X: 43) were included. Patient/operative factors were similar between robot systems ( P > .05). The pedicle screw accuracy was similar between robots (Renaissance: 1.1%% vs. X: 1.3%, P = .786); however, the S2AI screw breach rate was significantly lower for the X (Renaissance: 9.5% vs. X: 1.2%, P = .025). Robot time per screw was not statistically different (Renaissance: 4.6 minutes vs. X: 3.9 minutes, P = .246). The X was more reliable with an abandonment rate of 2.3% vs. Renaissance:22.7%, P = .007. Radiation exposure were not different between robot systems. Non-robot related complications including dural tear, loss of motor/sensory function, and blood transfusion were similar between robot systems. Conclusion: This is the first comparative analyses of screw accuracy, robot time/screw, robot abandonment, and radiation exposure between the Mazor X and Renaissance systems. There are substantial improvements in the X robot, particularly in the perioperative planning processes, which likely contribute to the X’s superiority in S2AI screw accuracy by nearly 8-fold and robot reliability by nearly 10-fold.
- Subjects :
- 030222 orthopedics
medicine.medical_specialty
Adult patients
business.industry
The Renaissance
Radiation exposure
03 medical and health sciences
0302 clinical medicine
Robotic systems
Physical medicine and rehabilitation
Propensity score matching
Abandonment (emotional)
medicine
Robot
Orthopedics and Sports Medicine
Surgery
Neurology (clinical)
business
030217 neurology & neurosurgery
Subjects
Details
- ISSN :
- 21925682
- Database :
- OpenAIRE
- Journal :
- Global spine journal
- Accession number :
- edsair.doi.dedup.....35a57c5f7977c46663f64a30b4de3a97