Back to Search
Start Over
A single liquid chromatography-quadrupole time-of-flight mass spectrometric method for the quantification of total antibody, antibody-conjugated drug and free payload of antibody-drug conjugates
- Source :
- Biomedical chromatography : BMC. 32(7)
- Publication Year :
- 2017
-
Abstract
- A single hybrid affinity-captured-LC-TOF-MS/MS method was developed and applied for the quantification of total antibody, antibody conjugated drug and free payload of antibody drug conjugate (ADC). Adcetris®, a valine-citrulline monomethyl auristatin E conjugated ADC, was used as a model ADC compound. A quadratic regression (weighted 1/concentration) was used to fit calibration curves over the concentration range 30.65-613.00 ng/mL with an equation y = ax2 + bx + c for the antibody-conjugated drug of Adcetris®. The qualification run met the acceptance criteria of ±25% accuracy and precision values for quality control samples. For the analysis of total antibody, a signature peptide (TTPPVLDSDGSFFLYSK, molecular weight 1874) was used after affinity capture using magnetic beads and on-bead trypsin digestion. A quadratic regression (weighted 1/concentration) was used to fit calibration curves over the concentration range 5.00-100.00 μg/mL with an equation y = ax2 + bx + c for total antibody. For free payload analysis of monomethyl auristatin E, a protein precipitation method followed by LC-TOF-MS/MS analysis was used. A quadratic regression (weighted 1/concentration) was used to fit calibration curves over the concentration range 1.01-2200 ng/mL with an equation y = ax2 + bx + c for free payload. Pharmacokinetic study samples and in vitro stability samples in rat were successfully analyzed by this a hybrid affinity-captured-LC-TOF-MS/MS method. This single platform method is a useful complementary method for the pharmacokinetics study of ADC with valine-citrulline linker at the early drug discovery stage.
- Subjects :
- Accuracy and precision
Antibody-drug conjugate
Immunoconjugates
Calibration curve
Clinical Biochemistry
030226 pharmacology & pharmacy
01 natural sciences
Biochemistry
Sensitivity and Specificity
Analytical Chemistry
03 medical and health sciences
chemistry.chemical_compound
0302 clinical medicine
Pharmacokinetics
Tandem Mass Spectrometry
Drug Discovery
Animals
Molecular Biology
Pharmacology
Polynomial regression
Brentuximab Vedotin
Chromatography
Chemistry
Protein Stability
010401 analytical chemistry
Reproducibility of Results
General Medicine
0104 chemical sciences
Rats
Monomethyl auristatin E
Linker
Conjugate
Chromatography, Liquid
Subjects
Details
- ISSN :
- 10990801
- Volume :
- 32
- Issue :
- 7
- Database :
- OpenAIRE
- Journal :
- Biomedical chromatography : BMC
- Accession number :
- edsair.doi.dedup.....35648e02720d48c0837c11eebbd8fc14