Validation of a Paper Chromatographic Methodology as an Alternative for Determination of the Radiochemical Purity of Na18F

Authors :
J.O. Nicolini
Patricia D. Zubata
Marcela Zubillaga
Natalia M. Leonardi
María J. Salgueiro
Guillermo Arturo Casale
Source :
Journal of Nuclear Medicine Technology. 40:271-274
Publication Year :
2012
Publisher :
Society of Nuclear Medicine, 2012.

Abstract

The aim of the present work was to validate a paper chromatography system as an alternative way to determine the radiochemical purity of Na18F. Methods: The evaluated parameters were specificity, limit of quantification, measurement interval, linearity, precision, accuracy, and robustness. Results: The proposed method proved to be linear (P > 0.05; r2 = 1.000), precise (relative SD, 8.6%), accurate (mean recovery, 95.9%; relative SD, 1.5%–1.8%), and robust under different conditions since no influence of the operative variables on the chromatographic performance was observed. Conclusion: This system can be used as a reliable alternative method to determine the radiochemical purity of Na18F samples that can be easily performed in PET radiopharmacies at low cost.