Output characteristics of piezoelectric lead zirconate titanate detector using high-energy heavy-ion beam

Authors :
Masahiro Sekiguchi
Yukio Uchihori
Masayuki Fujii
Osamu Okudaira
Masanori Kobayashi
Takashi Miyachi
Seiji Takechi
Nagaya Okada
Maki Hattori
Hiromi Shibata
Takeshi Murakami
Source :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 737:52-55
Publication Year :
2014
Publisher :
Elsevier BV, 2014.

Abstract

A radiation detector fabricated using piezoelectric lead zirconate titanate (PZT) has been studied by irradiating it with a 400 MeV/n xenon (Xe) beam. The beam diameter was controlled to change the irradiation conditions. It was found that the magnitude of the output observed from the PZT detector may be related to the number of Xe ions per unit area per unit time within the limits of the experimental conditions.