Construction of 144, 565 keV and 5.0 MeV monoenergetic neutron calibration fields at JAERI

Authors :
S. Shimizu
Katsutoshi Fujii
Makoto Yoshida
Jun Saegusa
Y. Tanimura
Michio Yoshizawa
Akira Uritani
Y. Shibata
Katsuhisa Kudo
Source :
Radiation protection dosimetry. 110(1-4)
Publication Year :
2004

Abstract

Monoenergetic neutron calibration fields of 144, 565 keV and 5.0 MeV have been developed at the Facility of Radiation Standards of JAERI using a 4 MV Pelletron accelerator. The 7 Li(p,n) 7 Be and 2 H(d,n) 3 He reactions areemployed for neutron production. The neutron energy was measured by the time-of-flight method with a liquid scintillation detector and calculated with the MCNP-ANT code. A long counter is employed as a neutron monitor because of the flat response. The monitor is set up where the influence of inscattered neutrons from devices and their supporting materials at a calibration point is as small as possible. The calibration coefficients from the monitor counts to the neutron fluence at a calibration point were obtained from the reference fluence measured with the transfer instrument of the primary standard laboratory (AIST), a 24.13 cm Φ Bonner sphere counter. The traceability of the fields to AIST was established through the calibration.