Back to Search
Start Over
Construction of 144, 565 keV and 5.0 MeV monoenergetic neutron calibration fields at JAERI
- Source :
- Radiation protection dosimetry. 110(1-4)
- Publication Year :
- 2004
-
Abstract
- Monoenergetic neutron calibration fields of 144, 565 keV and 5.0 MeV have been developed at the Facility of Radiation Standards of JAERI using a 4 MV Pelletron accelerator. The 7 Li(p,n) 7 Be and 2 H(d,n) 3 He reactions areemployed for neutron production. The neutron energy was measured by the time-of-flight method with a liquid scintillation detector and calculated with the MCNP-ANT code. A long counter is employed as a neutron monitor because of the flat response. The monitor is set up where the influence of inscattered neutrons from devices and their supporting materials at a calibration point is as small as possible. The calibration coefficients from the monitor counts to the neutron fluence at a calibration point were obtained from the reference fluence measured with the transfer instrument of the primary standard laboratory (AIST), a 24.13 cm Φ Bonner sphere counter. The traceability of the fields to AIST was established through the calibration.
- Subjects :
- Astrophysics::High Energy Astrophysical Phenomena
Radiation Dosage
Nuclear physics
Fast Neutrons
Radiation Protection
Japan
Neutron flux
Calibration
Scattering, Radiation
Radiology, Nuclear Medicine and imaging
Neutron
Computer Simulation
Nuclear Experiment
Radiometry
Bonner sphere
Physics
Radioisotopes
Radiation
Neutron monitor
Radiological and Ultrasound Technology
Public Health, Environmental and Occupational Health
General Medicine
Equipment Design
Reference Standards
Neutron temperature
Pelletron
Equipment Failure Analysis
Models, Chemical
Primary standard
Monte Carlo Method
Algorithms
Subjects
Details
- ISSN :
- 01448420
- Volume :
- 110
- Issue :
- 1-4
- Database :
- OpenAIRE
- Journal :
- Radiation protection dosimetry
- Accession number :
- edsair.doi.dedup.....304fa14ce8591f42cb12acbff8ed3b43