Back to Search Start Over

Response of GAFChromic® HD-V2 film dosimeter in 10-300 Gy dose range for radiation testing of electronic devices

Authors :
Giuseppe Perrone
Aldo Parlato
Francesca Martorana
Elio Angelo Tomarchio
Martorana F
Parlato A
Perrone G
Tomarchio E
Source :
Nuclear Technology and Radiation Protection, Vol 34, Iss 1, Pp 79-84 (2019)
Publication Year :
2019
Publisher :
National Library of Serbia, 2019.

Abstract

The study reported in this paper aimed to investigate the response of a GAFChromic? HD-V2 film dosimeter in the dose range between 10 and 300 Gy, normally used to perform irradiation tests on electronic devices with a 60Co gamma-ray irradiator. The well-defined linearity of response in terms of absorbance as a function of absorbed dose, easiness of handling and data analysis of irradiated dosimeters, reproducibility, stability, and insensitivity to visible light and most of the environmental parameters, make HD-V2 film a flexible, inexpensive and reliable dose measurement device. The study has shown a fogging effect of the dosimeter response since its value changes over time. Strategies and adequate precautions to take into account changes on HD-V2 film physical properties for calibration and routine use in industrial applications, with a specific reference to radiation testing of electronic devices, have been discussed.