Reversal of flow during carotid artery stenting: use of the Parodi antiembolism system

Authors :
Tobias Struffert
Christoph Krick
F. Ahlhelm
Bernd F. M. Romaike
Wolfgang Reith
Iris Q. Grunwald
Panagiotis Papanagiotou
Maria Politi
Source :
Neuroradiology. 49(3)
Publication Year :
2006

Abstract

INTRODUCTION: To evaluate the flow reversal efficacy of the Parodi antiembolism system (PAES) in the prevention of distal emboli during carotid stenting. METHODS: A total of 90 patients were treated for internal carotid artery (ICA) stenosis with a stent device. A PAES device was used in 31 symptomatic and 5 asymptomatic patients (total 36 patients) with ICA stenosis, and 54 patients were stented without any protection device. Diffusion-weighted (DW) imaging was performed before and after stenting. RESULTS: In the group without PAES protection, 23 out of 54 patients showed new lesions on DW images after stenting. Of the lesions seen, 147 (2.72 lesions/patient) were in the vessel-dependent area. In the group with protection, 19 out of 36 patients had new lesions, and only 34 (0.94 lesions/patient) were noted in the vessel-dependent area. The number of new lesions in the nondependent vessel area did not differ if a protection system was used (P = 0.671). The use of PAES led to a significant reduction (P = 0.024) in the incidence of the most frequently seen lesions (size