Back to Search Start Over

Label-free sensing of the binding state of MUC1 peptide and anti-MUC1 aptamer solution in fluidic chip by terahertz spectroscopy

Authors :
Xiang Yang
Weiling Fu
Shihan Yan
Mengwan Liu
Ke Yang
Dongshan Wei
Hong-Liang Cui
Yunxia Wang
Xiang Zhao
Mingkun Zhang
Source :
Biomedical Optics Express. 8:4427
Publication Year :
2017
Publisher :
The Optical Society, 2017.

Abstract

The aptamer and target molecule binding reaction has been widely applied for construction of aptasensors, most of which are labeled methods. In contrast, terahertz technology proves to be a label-free sensing tool for biomedical applications. We utilize terahertz absorption spectroscopy and molecular dynamics simulation to investigate the variation of binding-induced collective vibration of hydrogen bond network in a mixed solution of MUC1 peptide and anti-MUC1 aptamer. The results show that binding-induced alterations of hydrogen bond numbers could be sensitively reflected by the variation of terahertz absorption coefficients of the mixed solution in a customized fluidic chip. The minimal detectable concentration is determined as 1 pmol/μL, which is approximately equal to the optimal immobilized concentration of aptasensors.

Details

ISSN :
21567085
Volume :
8
Database :
OpenAIRE
Journal :
Biomedical Optics Express
Accession number :
edsair.doi.dedup.....2e271f80e559081556a28e5c02270960
Full Text :
https://doi.org/10.1364/boe.8.004427