Abstract
- Background:Meniscal lesions represent one of the most common intra-articular knee injuries. Meniscus replacement devices are needed to restore load distribution and knee stability after meniscectomy. Fixation of these devices is crucial to the generation of hoop stresses and the distribution of loads in the joint.Purpose:To evaluate 2 different fixation techniques (suture endobutton and interference screw) for implantation of a novel meniscus device.Study Design:Controlled laboratory study.Methods:In 7 human cadaveric knees (aged 17-61 years), 1 anterior and 2 potential posterior tunnel locations were investigated, and both fixation techniques were tested in each tunnel. The native meniscus roots, devices fixed with a suture endobutton, and devices fixed with an interference screw were gripped with cryoclamps, and tibias were drilled and loaded into a custom jig. Samples were preloaded, preconditioned, loaded for 500 cycles (50-150 N), and tested in tension until failure.Results:For all 3 tunnels, suture fixation resulted in greater elongation (54.1%-150.7% greater; P < .05) during cyclic loading than interference screw fixation, which approximated the native roots. Both fixation techniques displayed ultimate tensile loads in the same range as native roots. However, stiffness of the suture fixation groups (36.5-41.6 N/mm) was only 28% to 37% of that of the interference screw fixation groups (98.7-131.6 N/mm), which had values approaching those of the native roots (anterior: 175.4 ± 24.2 N/mm; posterior: 157.6 ± 22.9 N/mm).Conclusion:Interference screw fixation was found to be superior to suture fixation with regard to elongation and stiffness, a finding that should be considered in the design and implantation of novel meniscus replacement devices.Clinical Relevance:With the emergence of various devices for total meniscus replacement, the establishment of fixation strategies is crucial for the generation of tensile hoop stresses and the efficacy of these approaches.
- Subjects :
- Adult
Adolescent
Knee Joint
Bone Screws
Physical Therapy, Sports Therapy and Rehabilitation
Load distribution
Knee Injuries
Young Adult
03 medical and health sciences
Fixation (surgical)
0302 clinical medicine
Cadaver
Humans
Medicine
Meniscus
Orthopedics and Sports Medicine
A fibers
Fibrous joint
Orthodontics
030222 orthopedics
Sutures
Tibia
business.industry
Biomechanics
030229 sport sciences
Middle Aged
Biomechanical Phenomena
medicine.anatomical_structure
business
Cadaveric spasm
Knee injuries
Meniscal lesions
Subjects
Details
- ISSN :
- 15523365 and 03635465
- Volume :
- 46
- Database :
- OpenAIRE
- Journal :
- The American Journal of Sports Medicine
- Accession number :
- edsair.doi.dedup.....2d9569cf6475b63c031bf7deaef1eeb0
- Full Text :
- https://doi.org/10.1177/0363546518773737