Back to Search
Start Over
Manual Fever Check Is More Sensitive than Infrared Thermoscanning Camera for Fever Screening in a Hospital Setting during the COVID-19 Pandemic
- Source :
- Journal of Korean Medical Science
- Publication Year :
- 2020
- Publisher :
- Korean Academy of Medical Sciences, 2020.
-
Abstract
- Between August 31st and September 4th 2020 in a tertiary care hospital in Seoul, the fever detection rate by infrared thermoscanning camera at the gates was 0.002% (95% confidence interval [CI], 0.001%–0.006%) and that by manual fever check at the outpatient clinics was 0.02% (95% CI, 0.01%–0.03%) (P < 0.001). The difference strongly suggests that sensitivity of the thermoscanning camera in the market must be upgraded.<br />Graphical Abstract
- Subjects :
- 2019-20 coronavirus outbreak
medicine.medical_specialty
Fever
Coronavirus disease 2019 (COVID-19)
Infrared Rays
Hospital setting
Pneumonia, Viral
Brief Communication
Sensitivity and Specificity
Infrared Thermometers
Betacoronavirus
03 medical and health sciences
0302 clinical medicine
Pandemic
medicine
Humans
Outpatient clinic
030212 general & internal medicine
Pandemics
SARS-CoV-2
business.industry
COVID-19
General Medicine
Tertiary care hospital
Hospitals
Confidence interval
Emergency medicine
Detection rate
Coronavirus Infections
business
Infrared Thermoscanner
Medicine General & Policy
Subjects
Details
- ISSN :
- 15986357 and 10118934
- Volume :
- 35
- Database :
- OpenAIRE
- Journal :
- Journal of Korean Medical Science
- Accession number :
- edsair.doi.dedup.....2bd1ec8fffe8ca625939ebc2bc130dd9
- Full Text :
- https://doi.org/10.3346/jkms.2020.35.e389