Development of achromatic full-field x-ray microscopy with compact imaging mirror system
- Source :
- Proceedings of SPIE - The International Society for Optical Engineering. 8851:885107
- Publication Year :
- 2013
- Publisher :
- The International Society for Optical Engineering, 2013.
-
Abstract
- Compact advanced Kirkpatrick–Baez optics are used to construct a microscope that is easy to align and robust against vibrations and thermal drifts. The entire length of the imaging mirror system is 286 mm, which is 34% shorter than the previous model. A spatial resolution test is performed in which magnified bright-field images of a pattern are taken with an X-ray camera at an energy of 10 keV at the BL29XUL beamline of SPring-8. A line-and-space pattern having a 50- nm width could be resolved, although the image contrast is low.
Details
- Language :
- English
- ISSN :
- 0277786X
- Volume :
- 8851
- Database :
- OpenAIRE
- Journal :
- Proceedings of SPIE - The International Society for Optical Engineering
- Accession number :