Accuracy of displayed values of tidal volume in the pediatric intensive care unit
- Source :
- Critical care medicine. 30(11)
- Publication Year :
- 2002
-
Abstract
- To assess the accuracy of the expired tidal volumes (VT(E)) displayed by one of the most frequently used ventilators that measures exhaled volume at the expiratory valve.Prospective study.The intensive care units of a pediatric tertiary referral center in London, UK.A total of 56 intubated children aged between 3 wks and 16.6 yrs who were clinically stable and ventilated with a Servo 300 ventilator.The CO2SMO Plus respiratory monitor, which measures flow at the airway opening, was validated using calibrated syringes and appropriate tracheal tubes and connections. Simultaneous in vivo recordings of VT(E) from the Servo 300 and CO2SMO Plus were compared before (displayed Servo VT(E)) and after (effective Servo VT(E)) compensating for ventilator circuit compliance.The in vitro accuracy of the CO2SMO Plus was within +/-5% over a wide range of volumes and measurement conditions. The displayed Servo 300 VT(E) overestimated the true VT(E) by between 2% and 91%. The magnitude of error varied within and between children, according to pressure change (peak inspiratory pressure minus positive end-expiratory pressure), VT(E), and circuit size. Mean (sd) error was 32% (20%) in 40 children with displayed Servo VT(E) of160 mL and 18% (6%) in 16 subjects with displayed Servo VT(E) of/=160 mL. After correcting for gas compression, effective VT(E) from the Servo 300 underestimated the true VT(E) by up to 64% in the smallest infants but continued to overestimate by as much as 29% in older children.The accuracy of tidal volume values is crucially dependent on the site of measurement. Unless measured at the airway opening, displayed values are an inconsistent and misleading indicator of the true volumes delivered.
- Subjects :
- medicine.medical_specialty
Adolescent
medicine.medical_treatment
Critical Care and Intensive Care Medicine
Intensive Care Units, Pediatric
law.invention
law
Intensive care
medicine
Intubation, Intratracheal
Tidal Volume
Humans
Prospective Studies
Intensive care medicine
Child
Tidal volume
Endotracheal tube
Monitoring, Physiologic
Mechanical ventilation
Pediatric intensive care unit
Pulmonary mechanics
business.industry
Infant, Newborn
Infant
Reproducibility of Results
Intensive care unit
Respiration, Artificial
Respiratory Function Tests
Child, Preschool
Linear Models
Referral center
business
Subjects
Details
- ISSN :
- 00903493
- Volume :
- 30
- Issue :
- 11
- Database :
- OpenAIRE
- Journal :
- Critical care medicine
- Accession number :
- edsair.doi.dedup.....2b3101b8302222f6d8ff791039e4c994