Calculation of absorbed dose distributions from dynamic wedges

Authors :
Anna Samuelsson
Pascal R.M. Storchi
Karl-Axel Johansson
Philip Verlinde
Evert Woudstra
Radiation Oncology
Source :
Physics in Medicine and Biology, 43, 1497-1506. IOP Publishing Ltd.
Publication Year :
1998

Abstract

In radiotherapy with photon beams, the use of dynamic wedges, which are obtained by the movement of one of the jaws, offers an increasing flexibility relative to the traditional use of metal wedges. But it is a disadvantage for the measurement of absorbed dose distributions, because the absorbed dose at each measurement point can only be obtained after a complete movement of the jaw. Consequently, for radiotherapy planning, an algorithm should be available that does not require measurements for any specific dynamically wedged beam, but is based on only a modest number of measurements. In this paper, an algorithm for the calculation of the dose distribution from dynamic wedges is described. This algorithm uses the convolution of pencil beam kernels with a non-uniform field function. These pencil beam kernels are derived from empirical data resulting from measurements of the open beam only.

Details

ISSN :
00319155
Volume :
43
Database :
OpenAIRE
Journal :
Physics in Medicine and Biology
Accession number :
edsair.doi.dedup.....29abaa3927faf83a7fe485bf8409ac56
Full Text :
https://doi.org/10.1088/0031-9155/43/6/009