Back to Search
Start Over
T1-Based Synthetic Magnetic Resonance Contrasts Improve Multiple Sclerosis and Focal Epilepsy Imaging at 7 T
- Source :
- Investigative Radiology, Investigative Radiology, Lippincott, Williams & Wilkins, 2021, 56 (2), pp.127-133. ⟨10.1097/RLI.0000000000000718⟩, Investigative Radiology, 2021, 56 (2), pp.127-133. ⟨10.1097/RLI.0000000000000718⟩
- Publication Year :
- 2020
- Publisher :
- Ovid Technologies (Wolters Kluwer Health), 2020.
-
Abstract
- Objectives<br />Ultra-high field magnetic resonance imaging (MRI) (>= 7 T) is a unique opportunity to improve the clinical diagnosis of brain pathologies, such as multiple sclerosis or focal epilepsy. However, several shortcomings of 7 T MRI, such as radiofrequency field inhomogeneities, could degrade image quality and hinder radiological interpretation. To address these challenges, an original synthetic MRI method based on T-1 mapping achieved with the magnetization-prepared 2 rapid acquisition gradient echo (MP2RAGE) sequence was developed. The radiological quality of on-demand T-1-based contrasts generated by this technique was evaluated in multiple sclerosis and focal epilepsy imaging at 7 T. Materials and Methods<br />This retrospective study was carried out from October 2017 to September 2019 and included 21 patients with different phenotypes of multiple sclerosis and 35 patients with focal epilepsy who underwent MRI brain examinations using a whole-body investigative 7 T magnetic resonance system. The quality of 2 proposed synthetic contrast images were assessed and compared with conventional images acquired at 7 T using the MP2RAGE sequence by 4 radiologists, evaluating 3 qualitative criteria: signal homogeneity, contrast intensity, and lesion visualization. Statistical analyses were performed on reported quality scores using Wilcoxon rank tests and further multiple comparisons tests. Intraobserver and interobserver reliabilities were calculated as well. Results<br />Radiological quality scores were reported higher for synthetic images when compared with original images, regardless of contrast, pathologies, or raters considered, with significant differences found for all 3 criteria (P < 0.0001, Wilcoxon rank test). None of the 4 radiologists ever rated a synthetic image "markedly worse" than an original image. Synthetic images were rated slightly less satisfying for only 3 epileptic patients, without precluding lesion identification. Conclusion<br />T-1-based synthetic MRI with the MP2RAGE sequence provided on-demand contrasts and high-quality images to the radiologist, facilitating lesion visualization in multiple sclerosis and focal epilepsy, while reducing the magnetic resonance examination total duration by removing an additional sequence.
- Subjects :
- medicine.medical_specialty
Magnetic Resonance Spectroscopy
genetic structures
Lesion Identification
Wilcoxon signed-rank test
Image quality
[SDV]Life Sciences [q-bio]
7 t
ultra-high field mri
multiple sclerosis
030218 nuclear medicine & medical imaging
Lesion
[SPI]Engineering Sciences [physics]
03 medical and health sciences
Epilepsy
0302 clinical medicine
medicine
Humans
neuroradiology
Radiology, Nuclear Medicine and imaging
ComputingMilieux_MISCELLANEOUS
Retrospective Studies
medicine.diagnostic_test
synthetic mri
business.industry
Multiple sclerosis
Brain
Magnetic resonance imaging
General Medicine
medicine.disease
Magnetic Resonance Imaging
3. Good health
Multiple comparisons problem
epilepsy
Epilepsies, Partial
Radiology
medicine.symptom
business
mp2rage
030217 neurology & neurosurgery
Subjects
Details
- ISSN :
- 15360210 and 00209996
- Volume :
- 56
- Database :
- OpenAIRE
- Journal :
- Investigative Radiology
- Accession number :
- edsair.doi.dedup.....275b062cba8a06c0a3a24c6595a6cd44
- Full Text :
- https://doi.org/10.1097/rli.0000000000000718