Back to Search Start Over

Prosthetic hip failure: retrospective radiograph image analysis of the acetabular cup

Authors :
D.W.L. Hukins
P.R. Jones
Kevin Hardinge
Martyn L. Porter
Christopher J. Taylor
Source :
University of Manchester-PURE
Publication Year :
1989

Abstract

A method has been developed for quantifying movement and wear of the acetabular component (cup) of total hip replacements (THR) from routine postoperative and review radiographs. The method uses both interactive and automatic computer image analysis techniques. Dimensions of the prosthesis are used to scale the measurements and so overcome variation in radiographic alignment. The application of the method is illustrated by retrospective investigations of cup migration and wear using review radiographs taken over a follow-up of at least 12 years.

Details

ISSN :
01415425
Volume :
11
Issue :
3
Database :
OpenAIRE
Journal :
Journal of biomedical engineering
Accession number :
edsair.doi.dedup.....25a83a9ac8fa25f446892f41c062db5d