Evaluation of beam hardening and photon scatter by brass compensator for IMRT
- Source :
- Journal of Radiation Research
- Publication Year :
- 2012
- Publisher :
- Oxford University Press, 2012.
-
Abstract
- When a brass compensator is set in a treatment beam, beam hardening may take place. This variation of the energy spectrum may affect the accuracy of dose calculation by a treatment planning system and the results of dose measurement of brass compensator intensity modulated radiation therapy (IMRT). In addition, when X-rays pass the compensator, scattered photons are generated within the compensator. Scattered photons may affect the monitor unit (MU) calculation. In this study, to evaluate the variation of dose distribution by the compensator, dose distribution was measured and energy spectrum was simulated using the Monte Carlo method. To investigate the influence of beam hardening for dose measurement using an ionization chamber, the beam quality correction factor was determined. Moreover, to clarify the effect of scattered photons generated within the compensator for the MU calculation, the head scatter factor was measured and energy spectrum analyses were performed. As a result, when X-rays passed the brass compensator, beam hardening occurred and dose distribution was varied. The variation of dose distribution and energy spectrum was larger with decreasing field size. This means that energy spectrum should be reproduced correctly to obtain high accuracy of dose calculation for the compensator IMRT. On the other hand, the influence of beam hardening on k(Q) was insignificant. Furthermore, scattered photons were generated within the compensator, and scattered photons affect the head scatter factor. These results show that scattered photons must be taken into account for MU calculation for brass compensator IMRT.
- Subjects :
- Technology
Photon
Health, Toxicology and Mutagenesis
Monte Carlo method
Physics::Medical Physics
Brass
Optics
scatter photon
Scattering, Radiation
Radiology, Nuclear Medicine and imaging
Computer Simulation
Physics
Monitor unit
Brass compensator IMRT
Photons
Radiation
Models, Statistical
business.industry
quality correction factor
Equipment Design
Equipment Failure Analysis
Zinc
beam hardening
visual_art
Ionization chamber
Beam hardening
visual_art.visual_art_medium
Laser beam quality
Radiotherapy, Conformal
business
dose distribution
Monte Carlo Method
Beam (structure)
Copper
Filtration
Subjects
Details
- Language :
- English
- ISSN :
- 13499157 and 04493060
- Volume :
- 53
- Issue :
- 6
- Database :
- OpenAIRE
- Journal :
- Journal of Radiation Research
- Accession number :
- edsair.doi.dedup.....24d6cfa8870389c2deeb235caede3e8c