Back to Search Start Over

Comparison of knee joint orientation in clinically versus biomechanically aligned computed tomography coordinate system

Authors :
Andrej M. Nowakowski
Thomas P. Scherer
Magdalena Müller-Gerbl
Sebastian Hoechel
Source :
Journal of Orthopaedic Translation, Vol 16, Iss, Pp 78-84 (2019), Journal of Orthopaedic Translation
Publication Year :
2019
Publisher :
Elsevier, 2019.

Abstract

Background: Preoperative planning of total knee arthroplasty is usually performed using knee-centred computed tomography (CT) data sets. The disadvantage of these data sets is having no account of the biomechanical axis of the lower extremity, known as Mikulicz line. It aligns the femoral head to the middle of the talocrural joint. For optimal prosthesis arrangement, the knee CT data set must therefore be brought in congruency with this line of loading to achieve the best results and eliminate rotational malalignments.This study aims to establish a relation between the knee-centred clinical coordinate system (CCS) and a biomechanical coordinate system (BCS) based on the Mikulicz line. Methods: CT data sets of 45 lower extremities were evaluated. Using VG Studio Max, a visualisation and measurement software program; each CT data set was aligned according to the CCS and BCS. After superimposing both the aligned data sets, the deviations of both coordinate systems in all three planes were measured with the centre of the knee defined as the origin. Results: For the coronal plane, the CCS was demonstrated to be 2.54° in adduction compared to the BCS [standard deviation (SD) = 1.8°]. In sagittal view, the CCS was demonstrated to be 0.3° retroversed (SD = 3.27°). Finally, the deviation in the axial plane showed an outward rotation of 3.39° (SD = 1.99°). The alignment as well as the measurements demonstrated high intraobserver and interobserver reproducibility. Conclusion: Both coordinate systems can be established in knee-centred CT data sets in a reproducible manner. Clearly, the CCS differs significantly from the BCS describing the biomechanical axis, but mathematical-based adaptations and corrections can be performed. The translational potential of this article: The findings of this study allow a mathematical conversion of a knee CT to the biomechanical axis of the leg. Keywords: Alignment, Computed tomography, Coordinate system, Mikulicz line, Total knee arthroplasty

Details

Language :
English
Volume :
16
Database :
OpenAIRE
Journal :
Journal of Orthopaedic Translation
Accession number :
edsair.doi.dedup.....2461d378f4935328a2d56f91b9653c60