Back to Search Start Over

Sizing Response of the Ultra-High Sensitivity Aerosol Size Spectrometer (UHSAS) and Laser Aerosol Spectrometer (LAS) to Changes in Submicron Aerosol Composition and Refractive Index

Authors :
Richard H. Moore
Elizabeth B. Wiggins
Adam T. Ahern
Stephen Zimmerman
Lauren Montgomery
Pedro Campuzano Jost
Claire E. Robinson
Luke D. Ziemba
Edward L. Winstead
Bruce E. Anderson
Charles A. Brock
Matthew D. Brown
Gao Chen
Ewan C. Crosbie
Hongyu Guo
Jose L. Jimenez
Carolyn E. Jordan
Ming Lyu
Benjamin A. Nault
Nicholas E. Rothfuss
Kevin J. Sanchez
Melinda Schueneman
Taylor J. Shingler
Michael A. Shook
Kenneth L. Thornhill
Nicholas L. Wagner
Jian Wang
Publication Year :
2021
Publisher :
Copernicus GmbH, 2021.

Abstract

We evaluate the sensitivity of the size calibrations of two commercially-available, high-resolution optical particle sizers to changes in aerosol composition and complex refractive index (RI). The Droplet Measurement Technologies Ultra-High Sensitivity Aerosol Size Spectrometer (UHSAS) and the TSI, Inc. Laser Aerosol Spectrometer (LAS) are two commonly used instruments for measuring the portion of the aerosol size distribution with diameters larger than nominally 60–90 nm. Both instruments illuminate particles with a laser and relate the single-particle light scattering intensity and count rate measured over a wide range of angles to the size-dependent particle concentration. While the optical block geometry and flow system are similar for each instrument, a significant difference between the two models is the laser wavelength (1054 nm for the UHSAS and 633 nm for the LAS) and intensity (about 100x higher for the UHSAS), which may affect the way each instrument sizes non-spherical or absorbing aerosols. Here, we challenge the UHSAS and LAS with laboratory-generated, mobility-size-classified aerosols of known chemical composition to quantify changes in the optical size response relative to that of ammonium sulphate (RI of 1.52 + 0i at 532 nm) and NIST-traceable polystyrene latex spheres (PSLs with RI of 1.59 + 0i at 589 nm). Aerosol inorganic salt species are chosen to cover the real refractive index range of 1.32 to 1.78, while chosen light-absorbing carbonaceous aerosols include fullerene soot, nigrosine dye, humic acid, and fulvic acid standards. The instrument response is generally in good agreement with the electrical mobility diameter. However, large undersizing deviations are observed for the low-refractive-index fluoride salts and the strongly absorbing nigrosine dye and fullerene soot particles. Polydisperse size distributions for both fresh and aged wildfire smoke aerosols from the recent Fire Influence on Regional to Global Environments Experiment – Air Quality (FIREX-AQ) and the Cloud, Aerosol, and Monsoon Processes Philippines Experiment (CAMP2EX) airborne campaigns show good agreement between both optical sizers and contemporaneous electrical mobility sizing and particle time-of-flight mass spectrometric measurements. We assess the instrument uncertainties by interpolating the laboratory response curves using previously-reported RIs and size distributions for multiple aerosol type classifications. These results suggest that, while the optical sizers may underperform for strongly absorbing laboratory compounds and fresh tailpipe emissions measurements, sampling aerosols within the atmospherically-relevant range of refractive indices are likely to be sized to better than ±10–20 % uncertainty over the submicron aerosol size range when using instruments calibrated with ammonium sulphate.

Subjects

Subjects :
complex mixtures

Details

ISSN :
18678548
Database :
OpenAIRE
Accession number :
edsair.doi.dedup.....2083ab5d6c74806a23213bff08bf9951