Interaction between infusion equipment resulting in drug overdose in a critically ill patient
- Source :
- Anaesthesia. 60:88-91
- Publication Year :
- 2005
- Publisher :
- Wiley, 2005.
-
Abstract
- A critically ill septic patient on haemofiltration for acute renal failure suffered sudden circulatory and respiratory collapse. The cause of the collapse was traced to an interaction between mechanical devices (syringe driver infusion pumps and haemofiltration equipment) connected to two central venous catheters.
- Subjects :
- medicine.medical_specialty
Adolescent
Critical Illness
Remifentanil
Drug overdose
Piperidines
medicine
Humans
Hypnotics and Sedatives
Intensive care medicine
Propofol
Infusion Pumps
Collapse (medical)
Syringe driver
business.industry
Critically ill
Acute Kidney Injury
medicine.disease
Analgesics, Opioid
Equipment failure
Anesthesiology and Pain Medicine
Anesthesia
Equipment Failure
Female
Drug Overdose
Hemofiltration
medicine.symptom
business
Mechanical devices