- Purpose: Peripheral nerve stimulation (PNS) modeling has a potential role in designing and operating MRI gradient coils but requires computationally demanding simulations of electromagnetic fields and neural responses. We demonstrate compression of an electromagnetic and neurodynamic model into a single versatile PNS matrix (P-matrix) defined on an intermediary Huygens’ surface to allow fast PNS characterization of arbitrary coil geometries and body positions. Methods: The Huygens’ surface approach divides PNS prediction into an extensive precomputation phase of the electromagnetic and neurodynamic responses, which is independent of coil geometry and patient position, and a fast coil-specific linear projection step connecting this information to a specific coil geometry. We validate the Huygens’ approach by performing PNS characterizations for 21 body and head gradients and comparing them with full electromagnetic-neurodynamic modeling. We demonstrate the value of Huygens’ surface-based PNS modeling by characterizing PNS-optimized coil windings for a wide range of patient positions and poses in two body models. Results: The PNS prediction using the Huygens’ P-matrix takes less than a minute (instead of hours to days) without compromising numerical accuracy (error ≤ 0.1%) compared to the full simulation. Using this tool, we demonstrate that coils optimized for PNS at the brain landmark using a male model can also improve PNS for other imaging applications (cardiac, abdominal, pelvic, and knee imaging) in both male and female models. Conclusion: Representing PNS information on a Huygens’ surface extended the approach’s ability to assess PNS across body positions and models and test the robustness of PNS optimization in gradient design.
- Subjects :
- Male
Surface (mathematics)
Electromagnetic field
gradient coil design
animal structures
Computer science
Peripheral nerve stimulation
Phase (waves)
MRI safety
Article
Matrix (mathematics)
Electromagnetic Fields
Robustness (computer science)
Position (vector)
peripheral nerve stimulation
Humans
neurodynamic nerve model
Computer Simulation
Radiology, Nuclear Medicine and imaging
Peripheral Nerves
magneto-stimulation thresholds
Brain
Magnetic Resonance Imaging
nervous system
Electromagnetic coil
electromagnetic field simulation
Female
Algorithm
Subjects
Details
- ISSN :
- 15222594 and 07403194
- Volume :
- 87
- Database :
- OpenAIRE
- Journal :
- Magnetic Resonance in Medicine
- Accession number :
- edsair.doi.dedup.....1e64a3e2ba672055d74cbd58dd80ecfa
- Full Text :
- https://doi.org/10.1002/mrm.28966