Effectiveness and safety of a new helmet CPAP configuration allowing tidal volume monitoring in patients with COVID-19
- Source :
- Pulmonology
- Publication Year :
- 2021
-
Abstract
- Background High generated tidal volumes (Vt) have been correlated with higher risk of self-induced lung injury and worse clinical outcome. This study aimed to evaluate the effectiveness and safety of a new helmet continuous positive airway pressure delivered (h-CPAP) configuration allowing Vt monitoring in patients affected by COVID-19. Methods This prospective observational study was performed in the respiratory intermediate care unit of University Hospital in Turin, Italy, between March 24th, and June 15th, 2020. Included patients were treated with CPAP via a single-limb intentional leak configuration by a turbine-driven ventilator, provided with a dedicated patch. Effectiveness and safety of the configuration and healthcare workers safety were the outcomes of the study. Main findings Thirty-five patients were included in this study. Median age was 67 years (IQR 57–76 years), and 30 patients (85.7%) were men. Median value of overall leaks (intentional plus unintentional) was 68 L/min (IQR 63–75). Reliability of Vt measurements was 100%. An out of scale of Vt (above 50% compared to the previous values) was never recorded. Six patients (17.1%) needed more than two helmet replacements, due to leak test >10 l/min. Arm oedema and skin breakdowns were reported in sixteen (45.7%) and seven (20%) patients respectively. Among the 63 healthcare workers involved in the care of COVID-19 patients during the study only one was positive at RT-PCR nasopharyngeal swab testing. Conclusions The use of h-CPAP for treating COVID-19 in this configuration allowed for reliable Vt monitoring. Further studies evaluating this configuration in larger patients’ cohorts are needed.
- Subjects :
- Pulmonary and Respiratory Medicine
Leak
Tidal volume monitoring
medicine.medical_specialty
Coronavirus disease 2019 (COVID-19)
List of abbreviations: ARF, Acute respiratory failure
medicine.medical_treatment
Lung injury
Tidal volume
Vt, Tidal volume
CPAP
Medicine
In patient
Continuous positive airway pressure
Helmet
CPAP, Continuous positive airway pressure
RICU, Respiratory intermediate care unit
business.industry
COVID-19
ICU, Intensive care unit
Emergency medicine
HFNT, High flow nasal therapy
Observational study
Original Article
TDVs, turbine-driven ventilators