Back to Search
Start Over
Comparison of Hemostatic Changes in Pump-driven Extracorporeal Carbon Dioxide Removal and Venovenous Extracorporeal Membrane Oxygenation
- Source :
- ASAIO Journal. 68:1407-1413
- Publication Year :
- 2022
- Publisher :
- Ovid Technologies (Wolters Kluwer Health), 2022.
-
Abstract
- Extracorporeal carbon dioxide removal (ECCO 2 R) has gained widespread use as a supposedly less invasive alternative for hypercapnic respiratory failure besides venovenous extracorporeal membrane oxygenation (VV ECMO). Despite technological advances, coagulation-related adverse events remain a major challenge in both therapies. The overlapping operating areas of VV ECMO and pump-driven ECCO 2 R could allow for a device selection targeted at the lowest risk of such complications. This retrospective analysis of 47 consecutive patients compared hemostatic changes between pump-driven ECCO 2 R (n = 23) and VV ECMO (n = 24) by application of linear mixed effect models. A significant decrease in platelet count, increase in D-dimer levels, and decrease of fibrinogen levels were observed. However, except for fibrinogen, the type of extracorporeal support did not have a significant effect on the time course of these parameters. Our findings suggest that in terms of hemocompatibility, pump-driven ECCO 2 R is not significantly different from VV ECMO.
Details
- ISSN :
- 10582916
- Volume :
- 68
- Database :
- OpenAIRE
- Journal :
- ASAIO Journal
- Accession number :
- edsair.doi.dedup.....191f974adf586304fcae9727ce9c6987
- Full Text :
- https://doi.org/10.1097/mat.0000000000001675