Back to Search
Start Over
Gross Trunnion Failure in the Bipolar Hemiarthroplasty; Raising Concern about Short Trunnion: A Case Report
- Source :
- Hip & Pelvis
- Publication Year :
- 2021
- Publisher :
- Korean Hip Society, 2021.
-
Abstract
- There have been some reports of gross trunnion failure (GTF) in total hip arthroplasty. Here, we report a case of GTF 19 years after bipolar hemiarthroplasty using a 28-mm head with a 14/16 taper bore. Compared to other GTF reports, the current case had some unusual aspects: bipolar hemiarthroplasty, 28-mm head, relatively late-onset, and no apparent findings of metallosis though a severe one was evident. A Computed tomography scout view provided valuable information in evaluating polyethylene, metal head, and neck inside the bipolar cup. The current report suggests a need for concerned regarding short trunnion length which may be associated with GTF.
- Subjects :
- medicine.medical_treatment
Dentistry
Bipolar hemiarthroplasty
Computed tomography
Case Report
Arthroplasty
03 medical and health sciences
0302 clinical medicine
stomatognathic system
Prosthesis failure
Medicine
Metallosis
Orthopedics and Sports Medicine
030222 orthopedics
Hip
medicine.diagnostic_test
business.industry
Scout view
030229 sport sciences
medicine.disease
stomatognathic diseases
Trunnion
Surgery
business
Total hip arthroplasty
Subjects
Details
- Language :
- English
- ISSN :
- 22873279 and 22873260
- Volume :
- 33
- Issue :
- 1
- Database :
- OpenAIRE
- Journal :
- Hip & Pelvis
- Accession number :
- edsair.doi.dedup.....17950c374858386b4239a2b8365cf94f