Low-energy, single-pulse surface stimulation defibrillates large mammalian ventricles

Authors :
Edward J. Vigmond
Richard D. Walton
Jason D. Bayer
Olivier Bernus
Angel Moreno
Source :
Heart Rhythm. 19:308-317
Publication Year :
2022
Publisher :
Elsevier BV, 2022.

Abstract

Strong electric shocks are the gold standard for ventricular defibrillation but are associated with pain and tissue damage. We hypothesized that targeting the excitable gap (EG) of reentry with low-energy surface stimulation is a less damaging and painless alternative for ventricular defibrillation.The purpose of this study was to determine the conditions under which low-energy surface stimulation defibrillates large mammalian ventricles.Low-energy surface stimulation was delivered with five electrodes that were 7 cm long and placed 1-2 cm apart on the endocardial and epicardial surfaces of perfused pig left ventricle (LV). Rapid pacing (4 Hz) was used to induce reentry from a single electrode. A 2 ms defibrillation pulse ≤0.5 A was delivered from all electrodes with a varied time delay from the end of the induction protocol (0.1-5 seconds). Optical mapping was performed and arrhythmia dynamics analyzed. For mechanistic insight, simulations of the VF induction and defibrillation protocols were performed in silico with an LV model emulating the experimental conditions and electrodes placed 0.25-2 cm apart.In living LV, reentry was induced with varying complexity and dominant frequencies ranging between 3.5 to 6.2 Hz over 8 seconds postinitiation. Low-energy defibrillation was achieved with energy60 mJ and electrode separations up to 2 cm for less complex arrhythmia. In simulations, defibrillation consistently occurred when stimulation captured75% of the EG, which blocked reentry2.9 mm in front of the leading reentrant wavefront.Defibrillation with low-energy, single-pulse surface stimulation is feasible with energies below the human pain threshold (100 mJ). Optimal defibrillation occurs when arrhythmia complexity is minimal and electrodes capture75% of the EG.

Details

ISSN :
15475271
Volume :
19
Database :
OpenAIRE
Journal :
Heart Rhythm
Accession number :
edsair.doi.dedup.....149dfed9de086fb15e2c53d1a533689c
Full Text :
https://doi.org/10.1016/j.hrthm.2021.10.006