Back to Search Start Over

A simple tube adapter to expedite and automate thawing of viably frozen cells

Authors :
Margaret H. Beddall
Pratip K. Chattopadhyay
Kathy Foulds
Mario Roederer
Shing-Fen Kao
Source :
Journal of immunological methods. 439
Publication Year :
2016

Abstract

Although cryopreserved cell specimens are used throughout biomedical research, the process for thawing samples is labor-intensive and prone to error. Here we describe a small laboratory device that couples an uncapped vial of frozen cells to a conical tube containing warm cell culture media. The entire complex is loaded directly into a centrifuge; within 5min, cells are thawed and diluted out of toxic cryopreservation medium. The recovery and viability of cells are slightly reduced compared to the common (traditional) method. However, antigen-specific T-cell function is not affected. Since no technician time is required (beyond uncapping of vials), our device allows the parallel processing of as many samples as a centrifuge can hold (up to 96, in some models). Moreover, since the samples are not thawed manually in a water bath, the problems associated with technician-to-technician differences in sample handling are minimized, as is the potential for contamination. Importantly, the elimination of substantial labor involving subjective decisions standardizes this process and can reduce variability in results from cryopreserved specimens.

Details

ISSN :
18727905
Volume :
439
Database :
OpenAIRE
Journal :
Journal of immunological methods
Accession number :
edsair.doi.dedup.....144e150e16204fd6c0b3e87b5ce3f106