In vivo compatibility of Dynesys® spinal implants: a case series of five retrieved periprosthetic tissue samples and corresponding implants
- Source :
- Neukamp, Michal Sarah; Röder, Christoph; Veruva, S Y; MacDonald, D W; Kurtz, S M; Steinbeck, M J (2015). In vivo compatibility of Dynesys(®) spinal implants: a case series of five retrieved periprosthetic tissue samples and corresponding implants. European spine journal, 24(5), pp. 1074-1084. Springer 10.1007/s00586-014-3705-0
- Publication Year :
- 2014
- Publisher :
- Springer Science and Business Media LLC, 2014.
-
Abstract
- PURPOSE To determine whether particulate debris is present in periprosthetic tissue from revised Dynesys(®) devices, and if present, elicits a biological tissue reaction. METHODS Five Dynesys(®) dynamic stabilization systems consisting of pedicle screws (Ti alloy), polycarbonate-urethane (PCU) spacers and a polyethylene-terephthalate (PET) cord were explanted for pain and screw loosening after a mean of 2.86 years (1.9-5.3 years). Optical microscopy and scanning electron microscopy were used to evaluate wear, deformation and surface damage, and attenuated total reflectance Fourier transform infrared spectroscopy to assess surface chemical composition of the spacers. Periprosthetic tissue morphology and wear debris were determined using light microscopy, and PCU and PET wear debris by polarized light microscopy. RESULTS All implants had surface damage on the PCU spacers consistent with scratches and plastic deformation; 3 of 5 exhibited abrasive wear zones. In addition to fraying of the outer fibers of the PET cords in five implants, one case also evidenced cord fracture. The pedicle screws were unremarkable. Patient periprosthetic tissues around the three implants with visible PCU damage contained wear debris and a corresponding macrophage infiltration. For the patient revised for cord fracture, the tissues also contained large wear particles (>10 μm) and giant cells. Tissues from the other two patients showed comparable morphologies consisting of dense fibrous tissue with no inflammation or wear debris. CONCLUSIONS This is the first study to evaluate wear accumulation and local tissue responses for explanted Dynesys(®) devices. Polymer wear debris and an associated foreign-body macrophage response were observed in three of five cases.
- Subjects :
- Adult
Giant Cells, Foreign-Body
Male
Dense connective tissue
medicine.medical_specialty
Scanning electron microscope
Wear debris
Periprosthetic
610 Medicine & health
law.invention
Optical microscope
Pedicle Screws
In vivo
law
Spectroscopy, Fourier Transform Infrared
Microscopy
Humans
Medicine
Orthopedics and Sports Medicine
Device Removal
Polarized light microscopy
Polycarboxylate Cement
Polyethylene Terephthalates
business.industry
Foreign-Body Reaction
Macrophages
Prostheses and Implants
Middle Aged
Spine
Prosthesis Failure
Surgery
Female
business
Biomedical engineering
Subjects
Details
- ISSN :
- 14320932 and 09406719
- Volume :
- 24
- Database :
- OpenAIRE
- Journal :
- European Spine Journal
- Accession number :
- edsair.doi.dedup.....1373bec27ceaa4936a6aa55f863947c9
- Full Text :
- https://doi.org/10.1007/s00586-014-3705-0