In vivo compatibility of Dynesys® spinal implants: a case series of five retrieved periprosthetic tissue samples and corresponding implants

Authors :
Christoph Roeder
Sai Y. Veruva
Michal Neukamp
Marla J. Steinbeck
Steven M. Kurtz
Daniel W. MacDonald
Source :
Neukamp, Michal Sarah; Röder, Christoph; Veruva, S Y; MacDonald, D W; Kurtz, S M; Steinbeck, M J (2015). In vivo compatibility of Dynesys(®) spinal implants: a case series of five retrieved periprosthetic tissue samples and corresponding implants. European spine journal, 24(5), pp. 1074-1084. Springer 10.1007/s00586-014-3705-0
Publication Year :
2014
Publisher :
Springer Science and Business Media LLC, 2014.

Abstract

PURPOSE To determine whether particulate debris is present in periprosthetic tissue from revised Dynesys(®) devices, and if present, elicits a biological tissue reaction. METHODS Five Dynesys(®) dynamic stabilization systems consisting of pedicle screws (Ti alloy), polycarbonate-urethane (PCU) spacers and a polyethylene-terephthalate (PET) cord were explanted for pain and screw loosening after a mean of 2.86 years (1.9-5.3 years). Optical microscopy and scanning electron microscopy were used to evaluate wear, deformation and surface damage, and attenuated total reflectance Fourier transform infrared spectroscopy to assess surface chemical composition of the spacers. Periprosthetic tissue morphology and wear debris were determined using light microscopy, and PCU and PET wear debris by polarized light microscopy. RESULTS All implants had surface damage on the PCU spacers consistent with scratches and plastic deformation; 3 of 5 exhibited abrasive wear zones. In addition to fraying of the outer fibers of the PET cords in five implants, one case also evidenced cord fracture. The pedicle screws were unremarkable. Patient periprosthetic tissues around the three implants with visible PCU damage contained wear debris and a corresponding macrophage infiltration. For the patient revised for cord fracture, the tissues also contained large wear particles (>10 μm) and giant cells. Tissues from the other two patients showed comparable morphologies consisting of dense fibrous tissue with no inflammation or wear debris. CONCLUSIONS This is the first study to evaluate wear accumulation and local tissue responses for explanted Dynesys(®) devices. Polymer wear debris and an associated foreign-body macrophage response were observed in three of five cases.

Details

ISSN :
14320932 and 09406719
Volume :
24
Database :
OpenAIRE
Journal :
European Spine Journal
Accession number :
edsair.doi.dedup.....1373bec27ceaa4936a6aa55f863947c9
Full Text :
https://doi.org/10.1007/s00586-014-3705-0