Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans
- Source :
- Bai, Chuanyong; Conwell, Richard; Kindem, Joel; Babla, Hetal; Gurley, Mike; De Los Santos, Romer; et al.(2010). Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans. Journal of Nuclear Cardiology, 17(3), pp 459-469. doi: 10.1007/s12350-010-9204-8. Retrieved from: http://www.escholarship.org/uc/item/3c94n8m6, Journal of Nuclear Cardiology
- Publication Year :
- 2010
- Publisher :
- eScholarship, University of California, 2010.
-
Abstract
- Background We developed a cardiac SPECT system (X-ACT) with low dose volume CT transmission-based attenuation correction (AC). Three solid-state detectors are configured to form a triple-head system for emission scans and reconfigured to form a 69-cm field-of-view detector arc for transmission scans. A near mono-energetic transmission line source is produced from the collimated fluorescence x-ray emitted from a lead target when the target is illuminated by a narrow polychromatic x-ray beam from an x-ray tube. Transmission scans can be completed in 1 min with insignificant patient dose (deep dose equivalent
- Subjects :
- Attenuation correction
Image quality
Cardiology
Image processing
Imaging phantom
Collimated light
Cardiac SPECT
Image Processing, Computer-Assisted
Medicine & Public Health
Medicine
Humans
Radiology, Nuclear Medicine and imaging
Tomography, Emission-Computed, Single-Photon
business.industry
Attenuation
Imaging / Radiology
Myocardial Perfusion Imaging
Heart
equipment and supplies
Solid-state detector
Radiology Nuclear Medicine and imaging
Attenuation coefficient
Fluorescence x-ray transmission source
Original Article
Tomography
Nuclear Medicine
Cardiology and Cardiovascular Medicine
business
Nuclear medicine
Correction for attenuation
Biomedical engineering
Subjects
Details
- Language :
- English
- Database :
- OpenAIRE
- Journal :
- Bai, Chuanyong; Conwell, Richard; Kindem, Joel; Babla, Hetal; Gurley, Mike; De Los Santos, Romer; et al.(2010). Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans. Journal of Nuclear Cardiology, 17(3), pp 459-469. doi: 10.1007/s12350-010-9204-8. Retrieved from: http://www.escholarship.org/uc/item/3c94n8m6, Journal of Nuclear Cardiology
- Accession number :
- edsair.doi.dedup.....1331aba4a61eea2fbf091b6add96da46
- Full Text :
- https://doi.org/10.1007/s12350-010-9204-8.