Back to Search Start Over

Variation in patient position and impact on carbon-ion scanning beam distribution during prostate treatment

Authors :
Shinichiro Mori
Y Shiraishi
Daigo Kuroiwa
Kouichi Shibayama
M Nakao
Hiroshi Tsuji
Minoru Tajiri
Taku Inaniwa
Katsuyuki Tanimoto
Kentaro Miki
Source :
The British journal of radiology. 88(1051)
Publication Year :
2015

Abstract

We assessed the impact of changes in patient position on carbon-ion scanning beam distribution during treatment for prostate cancer.68 patients were selected. Carbon-ion scanning dose was calculated. Two different planning target volumes (PTVs) were defined: PTV1 was the clinical target volume plus a set-up margin for the anterior/lateral sides and posterior side, while PTV2 was the same as PTV1 minus the posterior side. Total prescribed doses of 34.4 Gy [relative biological effectiveness (RBE)] and 17.2 Gy (RBE) were given to PTV1 and PTV2, respectively. To estimate the influence of geometric variations on dose distribution, the dose was recalculated on the rigidly shifted single planning CT based on two dimensional-three dimensional rigid registration of the orthogonal radiographs before and after treatment for the fraction of maximum positional changes.Intrafractional patient positional change values averaged over all patients throughout the treatment course were less than the target registration error = 2.00 mm and angular error = 1.27°. However, these maximum positional errors did not occur in all 12 treatment fractions. Even though large positional changes occurred during irradiation in all treatment fractions, lowest dose encompassing 95% of the target (D95)-PTV1 was98% of the prescribed dose.Intrafractional patient positional changes occurred during treatment beam irradiation and degraded carbon-ion beam dose distribution. Our evaluation did not consider non-rigid deformations, however, dose distribution was still within clinically acceptable levels.Inter- and intrafractional changes did not affect carbon-ion beam prostate treatment accuracy.

Details

ISSN :
1748880X
Volume :
88
Issue :
1051
Database :
OpenAIRE
Journal :
The British journal of radiology
Accession number :
edsair.doi.dedup.....132d9d56519ca21c6ba6f56728e384be