Back to Search Start Over

Evaluation of automatic probe atomization for determination of elements in urine and whole blood by electrothermal atomic absorption spectrometry

Authors :
David Littlejohn
Tariq Mahmood Ansari
Olubode O. Ajayi
Source :
Scopus-Elsevier
Publication Year :
1992
Publisher :
Royal Society of Chemistry (RSC), 1992.

Abstract

A procedure is described for the determination of Cd, Co, Cr, Mn, Ni and Pb in Seronorm and Lanonorm reference materials and clinical samples. Accurate analyses of diluted specimens were achieved against aqueous standards. The use of high atomization temperatures (2400–2750 °C) minimized chemical interference from matrix components. High temperatues also improved the characteristic mass of Ga for probe electrothermal atomic absorption spectrometry, but caused a reduction in the sensitivity of Pb due to increased diffusional loss effects. The relative standard deviation values reported for Cd in human urine (1.1–7.6 µg dm–3) were mostly

Details

ISSN :
13645544 and 02679477
Volume :
7
Database :
OpenAIRE
Journal :
Journal of Analytical Atomic Spectrometry
Accession number :
edsair.doi.dedup.....12c82850d93f94a204f1d7f0ab81e19c
Full Text :
https://doi.org/10.1039/ja9920700689