Back to Search Start Over

Evaluation of BD Phoenix and Microscan WalkAway for determination of fosfomycin susceptibility in Staphylococcus aureus

Authors :
Davide Ghibaudo
Anna Maria Barbui
Alessandro Bondi
Cristina Costa
Rossana Cavallo
Gabriele Bianco
Marco Peradotto
Source :
Diagnostic Microbiology and Infectious Disease. 99:115240
Publication Year :
2021
Publisher :
Elsevier BV, 2021.

Abstract

We evaluate the performance of BD Phoenix (BD Diagnostic; Hunt Valley, MD) and MiscroScan WalkAway (Beckman Coulter Inc; Carlsbad, CA) systems versus reference agar dilution method for fosfomycin susceptibility determination in 200 Staphylococcus aureus clinical isolates. Both methods exhibit ≤89.0% of categorica agreement and ≥63.3% of very major error. All commercial antimicrobial susceptibility systems show poor concordance with reference method.

Details

ISSN :
07328893
Volume :
99
Database :
OpenAIRE
Journal :
Diagnostic Microbiology and Infectious Disease
Accession number :