Back to Search
Start Over
Microbiological and clinical evaluation of the isolator lysis-centrifugation blood culture tube
- Publication Year :
- 1983
-
Abstract
- In a controlled evaluation of 6,010 blood cultures, the yield of clinically significant microorganisms was greater from a lysis-centrifugation system (Isolator, Du Pont Co.) than from a nonvented vacuum bottle containing tryptic soy broth with sodium polyanetholesulfonate and CO2 and a vented bottle containing biphasic brain heart infusion medium with sodium polyanetholesulfonate. The Isolator significantly increased the frequency of isolation of Staphylococcus aureus and Candida spp. and significantly decreased the time required for the detection of S. aureus, Pseudomonas aeruginosa, and Candida spp.; however, anaerobic bacteria were recovered significantly more frequently from nonvented bottles with tryptic soy broth, and pneumococci were recovered significantly more frequently from both bottle systems. Contamination of cultures was significantly greater with the Isolator system than with either bottle system. Regardless of the number of blood cultures obtained per septic episode, the Isolator detected microbiologically proven bacteremia or fungemia in a significantly greater number of patients and significantly decreased the time required for detection.
- Subjects :
- Microbiology (medical)
Adult
business.product_category
medicine.disease_cause
Tryptic soy broth
Microbiology
chemistry.chemical_compound
Sepsis
medicine
Bottle
Humans
Blood culture
Fungemia
medicine.diagnostic_test
Bacteria
business.industry
medicine.disease
Culture Media
Blood
chemistry
Staphylococcus aureus
Evaluation Studies as Topic
Bacteremia
Brain heart infusion
Anaerobic bacteria
business
Research Article
Subjects
Details
- Language :
- English
- Database :
- OpenAIRE
- Accession number :
- edsair.doi.dedup.....0a5ae2f3aced4ac982125c5f2293aba4