An accurate automated blood pressure device for use in pregnancy and pre-eclampsia: the Microlife 3BTO-A

Authors :
A Reinders
Alexandra Cuckson
J T M Lee
Andrew Shennan
Source :
BJOG : an international journal of obstetrics and gynaecology. 112(7)
Publication Year :
2005

Abstract

Objective To assess the accuracy of an automated blood pressure device (Microlife 3BTO-A) in pregnancyand pre-eclampsia according to the British Hypertension Society (BHS) protocol.Design Prospective observational study.Setting Antenatal ward and clinics at Guy’s and St Thomas’ Hospital, London, UK.Population One hundred and five pregnant women including 35 women with non-proteinuric hypertensionand 35 with pre-eclampsia.Methods Two trained observers took nine sequential same-arm measurements from each woman. Measurementsalternated between a mercury sphygmomanometer and the device.Main outcome measures Grading criteria of the BHS protocol (A/B grade ¼ pass; C/D grade ¼ fail).Results The device passed the BHS protocol by achieving an A/B grade. It also achieved criteria of theAssociation for the Advancement of Medical Instrumentation for systolic and diastolic pressuresrespectively, in normotensive [ 0.5 (5.7) mmHg; 0.07 (7.7) mmHg], non-proteinuric hypertensive[ 3.3 (6.9) mmHg; 2.4 (6.6) mmHg] and pre-eclamptic pregnancy [ 4.1 (6.4) mmHg; 1.3 (7.9) mmHg].Conclusion The Microlife 3BTO-A can be recommended for use in a pregnant population, including pre-eclampsia, according to the BHS protocol.INTRODUCTIONIntracranial haemorrhage is the leading cause of maternalmortality in women with pre-eclampsia.

Details

ISSN :
14700328
Volume :
112
Issue :
7
Database :
OpenAIRE
Journal :
BJOG : an international journal of obstetrics and gynaecology
Accession number :
edsair.doi.dedup.....0a520cc471f96c222d0b28e6aa1f4542