Evaluation of cumulative dose for cone-beam computed tomography (CBCT) scans within phantoms made from different compositions using Monte Carlo simulations

Authors :
Abdullah Abuhaimed
Colin J. Martin
David J Gentle
Kurian Oomen
Marimuthu Sankaralingam
Source :
Journal of Applied Clinical Medical Physics
Publication Year :
2015
Publisher :
Wiley, 2015.

Abstract

Measurement of cumulative dose f(0,150) with a small ionization chamber within standard polymethyl methacrylate (PMMA) CT head and body phantoms, 150 mm in length, is a possible practical method for cone‐beam computed tomography (CBCT) dosimetry. This differs from evaluating cumulative dose under scatter equilibrium conditions within an infinitely long phantom f(0,∞), which is proposed by AAPM TG‐111 for CBCT dosimetry. The aim of this study was to investigate the feasibility of using f(0,150) to estimate values for f(0,∞) in long head and body phantoms made of PMMA, polyethylene (PE), and water, using beam qualities for tube potentials of 80−140 kV. The study also investigated the possibility of using 150 mm PE phantoms for assessment of f(0,∞) within long PE phantoms, the ICRU/AAPM phantom. The influence of scan parameters, composition, and length of the phantoms was investigated. The capability of f(0,150) to assess f(0,∞) has been defined as the efficiency and assessed in terms of the ratios ϵ(f(0,150)/f(0,∞)). The efficiencies were calculated using Monte Carlo simulations for an On‐Board Imager (OBI) system mounted on a TrueBeam linear accelerator. Head and body scanning protocols with beams of width 40−500 mm were used. Efficiencies ϵ(PMMA/PMMA) and ϵ(PE/PE) as a function of beam width exhibited three separate regions. For beam widths

Details

ISSN :
15269914
Volume :
16
Database :
OpenAIRE
Journal :
Journal of Applied Clinical Medical Physics
Accession number :
edsair.doi.dedup.....0925e9cc947cc83fc77642f22fb62634