A method for the assessment of the output of irregularly shaped electron fields

Authors :
Peter Andre
John T. Washington
Douglas L. Jones
Mark D. Hafermann
Source :
The British Journal of Radiology. 63:59-64
Publication Year :
1990
Publisher :
British Institute of Radiology, 1990.

Abstract

A simple approach to the calculation of dose in cone-collimated electron fields is presented. The method accounts for variations in lateral scatter with field size using a Clarkson Integration. The reduction in output by an irregularly shaped aperture is evaluated based on measurements of the output at various distances from the aperture in a field 2.5 cm in diameter. The first 14 months of clinical use were analysed and it is shown that in 42 out of 600 calculations, the cutout factor was less than 0.95 and that in 77% of these, the calculation method was accurate to ±3%.