Back to Search Start Over

Comparison of three patent foramen ovale closure devices in a randomized trial (Amplatzer versus CardioSEAL-STARflex versus Helex occluder)

Authors :
Andreas Baranowski
Ralph Hein
Ilona Hoffmann
Franziska Büscheck
Nina Wunderlich
Horst Sievert
Corinna Heinisch
Nicolas Majunke
Margaret Taaffe
Michaela Leetz
Yves L Bayard
Madlen Reschke
Neil Wilson
Evelyn Fischer
Source :
The American journal of cardiology. 101(9)
Publication Year :
2007

Abstract

This randomized trial compared procedural complications and 30-day clinical outcomes of 3 patent foramen ovale (PFO) closure devices (Amplatzer, Helex, and CardioSEAL-STARflex). It examined 660 patients (361 men, 299 women, mean age 49.3 ± 1.9 years), with 220 patients per group. All patients had a history of paradoxical embolism. All PFO closures were successful technically. Exchange of devices for others was most frequently required for the Helex occluder (7 of 220) and 2 of 220 in either of the other groups. Three device embolizations in the Helex group were retrieved and replaced successfully. One patient with a Helex occluder developed a transient ischemic attack and recovered without treatment. A hemopericardium in that group was punctured without affecting the device. One tamponade in the Amplatzer group required surgical device explantation. In 8 of 660 patients in the CardioSEAL-STARflex group, thrombi resolved after anticoagulation. Sixteen patients (11 in the CardioSEAL-STARflex group, 3 in the Amplatzer group, and 2 in the Helex group) had episodes of atrial fibrillation. PFOs were closed completely in 143 of 220 patients (65%) in the Amplatzer group, 116 of 220 patients (52.7%) in the Helex group, and 137 of 220 patients (62.3%) in the CardioSEAL-STARflex group at 30 days with significant differences between the Helex and Amplatzer occluders (p = 0.0005) and the Helex and CardioSEAL-STARflex occluders (p = 0.0003). PFO closure can be performed safely with each device. In conclusion, the Helex occluder embolized more frequently. Device thrombus formation and paroxysmal atrial fibrillation were more common with the CardioSEAL-STARflex occluder.

Details

ISSN :
00029149
Volume :
101
Issue :
9
Database :
OpenAIRE
Journal :
The American journal of cardiology
Accession number :