SU-E-J-142: Gafchromic Film Dosimetry in Fluoroscopy for Dynamic Tumor Tracking Irradiation of the Lung Using XR-SP2 Model - A Phantom Study

Authors :
T. Mizowaki
Takehiro Shiinoki
Takahiro Fujimoto
Y. Ishihara
Masaki Kokubo
Y Matsuo
T. Nakai
A. Sawada
Y. Miyabe
M. Hiraoka
Source :
Medical physics. 39(6Part8)
Publication Year :
2017

Abstract

Purpose: We have recently developed a dynamic tumor tracking irradiation system using Vero4DRT (MHI‐Tm2000). It is needed to create a 4D correlation model between a fiducial marker implanted near a tumor and an external surrogate as a function of time by continuously acquiring both fluoroscopyimages and external surrogate signals. The purpose of this study was to propose a new dosimetry method using Gafchromic XR‐SP2 films to measuresurface dose by fluoroscopyimaging.Methods: First, half‐value layers (HVLs) were measured using aluminum(Al) thicknesses (15 mm) at 40125 kVp. Subsequently, several films were irradiated using various milliampere second values on a solid water phantom. The surface air kerma were also measured using the chamber to calculate the surface doses under the same condition. Then, the calibration curve of dose vs. pixel values was calculated. Finally, surface dose by fluoroscopyimaging was measured using several pieces of film taped on the chest phantom. Orthogonal X‐ray fluoroscopyimaging was simultaneously performed until completion of data acquisition for creating a 4D correlation model. Those films were scanned after irradiation using a flat‐bed scanner and converted to dose by calibration curve. Results: The HVLs for tube voltage within 40125 kVp ranged from 2.35 to 5.98 mm Al. The calibration curve between surface dose and pixel values was reasonably smooth. The differences between the measured and the calibrated doses were less than 3%. The hot spots with the maximum dose of 37.12 mGy were observed around the area overlapped by both fluoroscopic fields. Conclusions: We have proposed a new dosimetry method using Gafchromic XR‐SP2 films to measuresurface dose by fluoroscopyimaging. This phantom study has demonstrated that it may be feasible to assess surface dose to patients during dynamic tumor tracking irradiation in clinic with ease after further investigation. This research was supported by the Japan Society for the Promotion of Science (JSPS) through its Funding Program for World‐Leading Innovation R&D on Science and Technology (FIRST Program). Research sponsored in part by Mitsubishi Heavy Industries, Ltd.

Details

ISSN :
24734209
Volume :
39
Issue :
6Part8
Database :
OpenAIRE
Journal :
Medical physics
Accession number :
edsair.doi.dedup.....03b65ffd6eaf13798914cd062ffa9a4a