Back to Search
Start Over
Thrombus Formation Patterns in the HeartMate II Ventricular Assist Device
- Source :
- ASAIO Journal. 60:237-240
- Publication Year :
- 2014
- Publisher :
- Ovid Technologies (Wolters Kluwer Health), 2014.
-
Abstract
- Postimplant device thrombosis remains a life-threatening complication and limitation of continuous-flow ventricular assist devices (VADs). Using advanced computational fluid dynamic (CFD) simulations, we successfully depicted various flow patterns, recirculation zones, and stagnant platelet trajectories which promote thrombus formation and observed that they matched actual thrombus formation patterns observed in Thoratec HeartMate II VADs explanted from patients with pump thrombosis. Previously, these small eddies could not be captured by either digital particle image velocimetry or CFD due to insufficient resolution. Our study successfully demonstrated the potential capability of advanced CFD to be adopted for device optimization, leading to enhanced safety and efficacy of VADs for long-term destination therapy.
- Subjects :
- medicine.medical_specialty
Materials science
medicine.medical_treatment
Biomedical Engineering
Biophysics
Bioengineering
Article
Biomaterials
Internal medicine
medicine
Humans
cardiovascular diseases
Thrombus
Pump thrombosis
Heartmate ii
Extramural
Thrombosis
General Medicine
Models, Theoretical
Flow pattern
medicine.disease
Particle image velocimetry
Pulsatile Flow
Ventricular assist device
Hydrodynamics
Cardiology
Heart-Assist Devices
Biomedical engineering
Destination therapy
Subjects
Details
- ISSN :
- 10582916
- Volume :
- 60
- Database :
- OpenAIRE
- Journal :
- ASAIO Journal
- Accession number :
- edsair.doi.dedup.....0232b320af5a43654239a07480b84418