Diabetes mellitus: The long way of standardization of HbA1c to the level of highest metrological order

Authors :
Kaiser, Patricia
Reinauer, Hans
Source :
GMS German Medical Science, Vol 9, p Doc28 (2011), GMS German Medical Science; VOL: 9; DOC28 /20111109/, GMS German Medical Science
Publication Year :
2011
Publisher :
German Medical Science GMS Publishing House, 2011.

Abstract

Glycated haemoglobin (HbA1c) measurements are used in clinical studies and for the management of diabetic patients. Various efforts were made to standardize the HbA1c measurements with consensus standards and standards based on a reference measurement procedure with external calibration. According to ISO 17511 a standard should meet highest accuracy possible, have a defined uncertainty of measurement and the calibration should be traceable to SI units. For HbA1c this has been realized using a LC-ID-MS procedure based on the existing reference measurement procedure.<br />GMS German Medical Science; 9:Doc28; ISSN 1612-3174