Multi-Channel High-Tc SQUID Detection System for Metallic Contaminants

Authors :
Yoshimi Hatsukade
Suzuki Shuichi
Y. Kitamura
Saburo Tanaka
Takeyoshi Ohtani
Source :
Physics Procedia. :156-161
Publisher :
Published by Elsevier B.V.

Abstract

Finding ultra-small metallic contaminants is a big issue for manufacturers of lithium-ion batteries nowadays. Therefore, we have developed high- T c SQUID systems for detection of such fine magnetic metallic contaminants. In this paper, we constructed an eight channel high- Tc SQUID gradiometer system for inspection of a sheet electrode of a lithium ion battery with width of about 70 mm. By this system, a small iron ball of about 30 μm in diameter was successfully detected. It is shown that this system has a detectable range of 70 mm in width. These results suggest that the system is a promising tool for the detection of the contaminants in lithium ion batteries.

Details

Language :
English
ISSN :
18753892
Database :
OpenAIRE
Journal :
Physics Procedia
Accession number :
edsair.doi.dedup.....01d8f54b5a7f6c2ca02463683681520d
Full Text :
https://doi.org/10.1016/j.phpro.2012.06.063