Back to Search Start Over

AD8232 to Biopotentials Sensors: Open Source Project and Benchmark

Authors :
José Jair Alves Mendes Junior
Daniel Prado Campos
Lorenzo Coelho de Andrade Villela De Biassio
Pedro Carlin Passos
Paulo Broniera Júnior
André Eugênio Lazzaretti
Eddy Krueger
Source :
Electronics, Volume 12, Issue 4, Pages: 833
Publication Year :
2023
Publisher :
Multidisciplinary Digital Publishing Institute, 2023.

Abstract

Acquiring biopotentials with fidelity using low-cost circuits is a significant challenge in biomedical instrumentation. In this perspective, our goal is to investigate the characteristics of the widely applied AD8232®, an analog front-end for biopotential acquisition. We designed and evaluated circuits to acquire the most common biosignals: electrocardiogram (ECG), electromyogram (EMG), and electroencephalogram (EEG). Our findings show that the circuit is suitable for ECG and EMG instrumentation, although it has limitations for EEG signals, particularly concerning the gain. The entire project of the boards is also a contribution of this work as we intend to corroborate open-source do-it-yourself biomedical instrumentation.

Details

Language :
English
ISSN :
20799292
Database :
OpenAIRE
Journal :
Electronics
Accession number :
edsair.doi.dedup.....00a914d6ffb8533695c61d978b6d3fb4
Full Text :
https://doi.org/10.3390/electronics12040833