Back to Search Start Over

Point projection radiography with the flexible x-ray imager

Authors :
Ted Perry
D. Hargrove
A. Nikitin
J. R. Mazuch
S. A. Alvarez
K. S. Budil
Perry M. Bell
Source :
Review of Scientific Instruments. 68:796-798
Publication Year :
1997
Publisher :
AIP Publishing, 1997.

Abstract

Radiography techniques utilizing large-area x-ray sources (typically ⩽7 keV) and pinhole-imaging gated x-ray diagnostics have long been used at the Nova laser facility. However, for targets requiring higher-energy x-ray backlighters (>9 keV), low conversion efficiencies and pinhole losses combine to make this scheme unworkable. The technique of point projection radiography has been improved upon to make imaging at high x-ray energies feasible. In this scheme a “point” source of x-rays, usually a small diameter (⩽25 μm) fiber, is illuminated with a single, 100-ps pulse from the Nova laser. A gated x-ray imager with a 500-ps electronic gate width is used to record the projected image. [K. S. Budil et al., Rev. Sci. Instrum. 67, 485 (1996).] The experimental challenges this technique presents and experimental results will be discussed.

Details

ISSN :
10897623 and 00346748
Volume :
68
Database :
OpenAIRE
Journal :
Review of Scientific Instruments
Accession number :
edsair.doi...........f93583f96f75b0d26f321bb6f3419a20
Full Text :
https://doi.org/10.1063/1.1147744