Low noise scintillation detectors with a P-47 thin layer screen for electrons of several keV

Authors :
P. Griesbach
K. Hildebrand
M. A. C. Ligtenberg
R. Albrecht
W. Meisel
R. Kurz
Zs. Kajcsos
P. Gütlich
Ch. Sauer
Source :
Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 93:505-515
Publication Year :
1994
Publisher :
Elsevier BV, 1994.

Abstract

The applicability of a low-noise scintillation detector (ScD) for the registration of electrons of several keV energy has been studied employing photomultipliers (PM) of different types and sizes. With the application of a sedimented P-47 scintillation screen, the values of the low-energy sensitivity limit and those of the light conversion coefficient were determined as about 2.7–4.7 keV and 2.8–6.6 photoelectrons/keV, respectively, for the set of PM's (Philips-Valvo XP 2020, Philips-Valvo XP 2052, Philips-Valvo XP 2972, EMI 9124a) studied. It is concluded that such scintillation detectors might be used advantageously as electron counters in the range of E > 5 keV. Applications below this kinetic energy value are also feasible when applying a floating acceleration of several kV to the ScD — a voltage much lower than the values required for Everhart-Thornley detectors.