Optimization of exposure parameters in digital tomosynthesis considering effective dose and image quality

Authors :
Seungyeon Choi
Sunghoon Choi
Hee-Joung Kim
Ye-Seul Kim
Dong-Hyuk Jang
Pil-Hyun Jeon
Dong-Hoon Lee
Haenghwa Lee
Source :
SPIE Proceedings.
Publication Year :
2016
Publisher :
SPIE, 2016.

Abstract

Digital tomosynthesis system (DTS), which scans an object in a limited angle, has been considered as an innovative imaging modality which can present lower patient dose than computed tomography and solve the problem of poor depth resolution in conventional digital radiography. Although it has many powerful advantages, only breast tomosynthesis system has been adopted in many hospitals. In order to reduce the patient dose while maintaining image quality, the acquisition conditions need to be studied. In this study, we analyzed effective dose and image qualities of chest phantom using commercialized universal chest digital tomosynthesis (CDT) R/F system to study the optimized exposure parameters. We set 10 different acquisition conditions including the default acquisition condition by user manual of Shimadzu (100 kVp with 0.5 mAs). The effective dose was calculated from PCXMC software version 1.5.1 by utilizing the total X-ray exposure measured by ion chamber. The image quality was evaluated by signal difference to noise ratio (SDNR) in the regions of interest (ROIs) pulmonary arteries at different axial in-plane. We analyzed a figure of merit (FOM) which considers both the effective dose and the SDNR in order to determine the optimal acquisition condition. The results indicated that the most suitable acquisition parameters among 10 conditions were condition 7 and 8 (120 kVp with 0.04 mAs and 0.1 mAs, respectively), which indicated lower effective dose while maintaining reasonable SDNRs and FOMs for three specified regions. Further studies are needed to be conducted for detailed outcomes in CDT acquisition conditions.

Details

ISSN :
0277786X
Database :
OpenAIRE
Journal :
SPIE Proceedings
Accession number :
edsair.doi...........f27e5311ade6ebd75bba340883043b64