Back to Search Start Over

Analysis of spherical aberration of a water immersion objective: application to specimens with refractive indices 1.33-1.40

Authors :
Robert H. Webb
Milind Rajadhyaksha
D.-S. Wan
Source :
Journal of Microscopy. 197:274-284
Publication Year :
2000
Publisher :
Wiley, 2000.

Abstract

The method of using immersion medium to correct spherical aberration for water immersion objectives when the samples are not water is investigated. Spherical aberration is measured by an interferometer converted from a confocal microscope for samples with different refractive indices. When the proper refractive index of the immersion medium and thickness of cover slip are selected, the measured spherical aberration approaches zero. A theoretical model can be used for prediction of the immersion medium to correct spherical aberration for various samples. Using the thinnest available cover slip (100 microm), the zero spherical aberration condition can be applied to samples with refractive index as high as 1.40. Confocal images in the condition of almost no spherical aberration are included to demonstrate the improvement of axial resolution due to this correction.

Details

ISSN :
00222720
Volume :
197
Database :
OpenAIRE
Journal :
Journal of Microscopy
Accession number :
edsair.doi...........ebb5158ce1676ac5419400f551b1b41d
Full Text :
https://doi.org/10.1046/j.1365-2818.2000.00635.x