Back to Search Start Over

Fast timing with plastic scintillators for in-beam heavy-ion spectroscopy

Authors :
H. J. Wollersheim
Dirk Rudolph
S. Pietri
Simon Emde
Robert Hoischen
N. Kurz
Anna Wendt
W. Prokopowicz
Pavel Golubev
H. Schaffner
J. Gerl
Source :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 654:354-360
Publication Year :
2011
Publisher :
Elsevier BV, 2011.

Abstract

The design, R&D, and testing of a new plastic-scintillator detector for Time-of-Flight measurements with relativistic heavy-ion beams are presented. A design approach using 32 independent precise timing measurements of the same physical event is followed. This is different from the conventional scheme, which aims at two or four high-precision measurements. A circular, 27 cm in diameter, BC-420 plastic-scintillator sheet is read-out by 32 photomultiplier tubes in order to achieve an intrinsic detector resolution on the order of 10 ps root mean square.