Back to Search Start Over

ECG gated circular cone-beam multi-cycle short-scan reconstruction algorithm

Authors :
Ralf Koester
Peter Koken
Gerhard Adam
Udo Van Stevendaal
P. G. Begemann
Michael Grass
Source :
SPIE Proceedings.
Publication Year :
2007
Publisher :
SPIE, 2007.

Abstract

In this contribution, the results of a phantom study for in-stent restenosis imaging with ECG gated continuous circular acquisition and reconstruction are summarized. Different rotation speeds and angular ranges are used to enable high resolution 3D and 4D reconstruction of objects covered by the cone at a high temporal resolution. Though the detector coverage of today's CT scanners is not large enough to irradiate the complete human heart, the coverage is sufficient to image smaller objects like conventional stents. We applied the proposed method to the visualization of an in-stent re-stenoses phantom covered by a clinical stent, attached to a dynamic heart phantom. The method delivers images of stents in vitro at an excellent visibility and is able to rule out in-stent occlusions.

Details

ISSN :
0277786X
Database :
OpenAIRE
Journal :
SPIE Proceedings
Accession number :
edsair.doi...........e67acfc913d9ed0c9d23267f2bb30123
Full Text :
https://doi.org/10.1117/12.707854